Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
3D navigation system allows remarkable reduction in fluoroscopy use during cavo-tricuspid isthmus ablation
Benoit Robaye1, Olivier Deceuninck2,3, Dominique Blommaert2
1Department of Cardiology, Université catholique de Louvain, CHU UCL Namur - Cliniques universitaire de Mont-Godinne, Av Dr G Therasse, 1, 5530, Yvoir, Belgium. benoit.robaye@uclouvain.be.
Purpose:
Typical atrial flutter (AFL) is one of the most common supraventricular arrhythmias. Its treatment mainly relies on cavo-tricuspid isthmus (CTI) ablation, which can be performed either using conventional fluoroscopy, still mainly used, or 3D navigation system to track the position of the catheter. The aim of this study is to show that the use of a 3D navigation system allows a dramatic reduction of fluoroscopy use during CTI ablation, without any loss of efficacy, time, or safety.
Methods:
In this single-center study, we retrospectively compared 134 cases of CTI ablation performed for typical AFL without a 3D navigation system with 95 cases of CTI ablation performed with such a 3D system. We compared the rates of procedural success (defined as obtaining a bidirectional electrical conduction block), freedom from AFL recurrence at 1-year follow-up, procedural time and safety, and fluoroscopy use.
Results:
Compared to conventional fluoroscopy, the use of a 3D navigation system significantly decreased the duration of fluoroscopy use (2 min 13 s ± 2 min 16 s versus 14 min 41 s ± 10 min 39 s, p < 0.0001) and dose-area products (1567.9 ± 1329.5 mGy cm2 versus 8263.3 ± 8636.6 mGy cm2, p < 0.0001). Procedure success rates, duration, and safety were not different between groups.
Conclusions:
The use of 3D navigation during CTI ablation substantially reduces fluoroscopy use duration, without reducing the success rates and safety or prolonging the procedure duration, as compared to conventional fluoroscopy. We therefore suggest the generalization of this navigation system.