Capacitive monitoring system for real-time respiratory motion monitoring during radiation therapy
Parisa Sadeghi1, Kathryn Moran2,3, James L Robar1,2
1Department of Physics and Atmospheric Science, Dalhousie University, Halifax, Nova Scotia, Canada.
Abstract:
This work introduces a novel capacitive-sensing technology capable of detecting respiratory motion with high temporal frequency (200 Hz). The system does not require contact with the patient and has the capacity to sense motion through clothing or plastic immobilization devices.
Abstract:
PURPOSE: This work presents and evaluates a novel capacitive monitoring system (CMS) technology for continuous detection of respiratory motion during radiation therapy. This modular system provides real-time motion monitoring without any contact with the patient, ionizing radiation, or surrogates such as reflective markers on the skin.
Materials And Methods:
The novel prototype features an array of capacitive detectors that are sensitive to the position of the body and capable of high temporal frequency readout. Performance of this system was investigated in comparison to the RPM infrared (IR) monitoring system (Varian Medical Systems). The prototype included three (5 cm × 10 cm) capacitive copper sensors in one plane, located at a distance of 8-10 cm from the volunteer. Capacitive measurements were acquired for central and lateral-to-central locations during chest free-breathing and abdominal breathing. The RPM IR data were acquired with the reflector block at corresponding positions simultaneously. The system was also tested during deep inspiration and expiration breath-hold maneuvers.
Results:
Capacitive monitoring system data demonstrate close agreement with the RPM status quo at all locations examined. Cross-correlation analysis on RPM and CMS data showed an average absolute lag of 0.07 s (range: 0.03-0.23 s) for DIBH and DEBH data and 0.15 s (range: 0-0.43 s) for free-breathing. Amplitude difference between the normalized CMS and RPM signal during chest and abdominal breathing was within 0.15 for 94.3% of the data points after synchronization. CMS performance was not affected when the subject was clothed.
Conclusion:
This novel technology permits sensing of both free-breathing and breath-hold respiratory motion. It provides data comparable to the RPM system but without the need for an IR tracking camera in the treatment room or use of reflective markers on the patient.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Acute Respiratory Failure-III
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...