A Semi-Automated and Reproducible Biological-Based Method to Quantify Calcium Deposition In Vitro
Published on: June 2, 2022
Albumin and calcium reference interval using healthy individuals and a data-mining approach
N Jassam1, A Luvai2, D Narayanan3
1Department of Clinical Biochemistry, Harrogate Foundation Trust, Harrogate, UK.
Harmonizing calcium reference intervals shows promise across UK platforms, but albumin intervals exhibit significant variation, challenging common reference ranges. Further investigation is needed for albumin standardization.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Analytical Chemistry
Background:
- Harmonization of reference intervals is recommended for analytes with established calibration and traceability.
- The UK Pathology Harmony initiative has recently harmonized reference intervals for calcium and albumin.
- This study aimed to determine reference intervals for calcium and albumin on common UK analytical platforms.
Purpose of the Study:
- To assess the feasibility of harmonizing reference intervals for calcium and albumin.
- To compare newly determined reference intervals with existing UK Pathology Harmony intervals.
- To evaluate the performance of different analytical platforms in relation to these harmonized intervals.
Main Methods:
- Recruited a prospective reference population (n=299) following IFCC CRIDL criteria.
- Collected a second, indirect population from 14 primary care settings.
- Measured analytes using Roche, Abbott, Beckman, and Siemens analytical platforms and methods.
Main Results:
- Calcium reference intervals largely agreed with UK Pathology Harmony intervals (2.2-2.6 mmol/L) across three of four platforms.
- BCG methods yielded higher reference limits for both calcium and albumin compared to BCP methods.
- Indirect study revealed age-related variations in reference intervals, up to 14% for older populations.
Conclusions:
- Calcium reference intervals demonstrate potential for harmonization across different analytical platforms.
- Albumin reference intervals showed substantial variation, indicating that a common reference interval is not currently supported.
- Analytical platform and method choice significantly impact reference intervals, particularly for albumin.
More Related Videos
Related Concept Videos
Skeleton and Calcium Homeostasis
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Serum Studies: Renal Function Tests
Bioequivalence Data: Statistical Interpretation