Case:
We report a 67-year-old gentleman who presented with a painful unstable knee. He had undergone a successful total knee arthroplasty 12 years earlier and was highly functional. He presented with a 10-month history of mild pain, instability, and gait alteration. During revision surgery, there was a loss of bony support, and a fractured femoral component was identified. He required constrained revision components for reconstruction and made an uneventful recovery.
Conclusions:
Fractured femoral components are rare complications of modern primary total knee arthroplasty. Loss of bony support in critical areas of high loading will inevitably lead to catastrophic component failure.