Related Experiment Video
Updated: Dec 15, 2025

Author Spotlight: Advancements in Multiplex Detection of Respiratory Viruses
Published on: November 10, 2023
Clinical evaluation of the BioFire® Respiratory Panel 2.1 and detection of SARS-CoV-2
Hannah M Creager1, Barbara Cabrera1, Andy Schnaubelt1
1University of Nebraska Medical Center, Omaha, NE, USA.
Abstract:
We evaluated the performance of the BioFire® Respiratory Panel 2.1 (RP2.1) in the detection of SARS CoV-2 in comparison against three other SARS CoV-2 EUA assays. In these studies, the RP2.1 panel had 98 % positive percent agreement (48/49) and 100 % negative percent agreement (49/49). Since 30 % of nasopharyngeal swab specimens have a SARS CoV-2 Ct >30 and thus detection of virus in low titers is clinically relevant, a sample with a high titer was diluted and each 10 fold dilution was tested in triplicate and compared against 6 other EUA approved SARS CoV-2 assays. These data suggested that the BioFire® RP2.1 panel, along with four other SARS CoV-2 assays (Roche cobas, Cepheid Xpert Xpress, BioFire® Defense COVID19, and NECoV19), consistently detected viral RNA at the 10-7 dilution. Overall, these studies suggest that the BioFire® RP2.1 assay can be used to detect acute cases of SARS CoV2 in addition to patients with low viral titer later in disease presentation.