Multicenter evaluation of Xpert HIV-1 viral load assay for HIV quantification in China
Lifeng Liu1,2, Renni Deng3, Yun Lan4
1Center for Infectious Diseases, Beijing Youan Hospital, Capital Medical University, Beijing, China.
Abstract:
New approaches to increase HIV-1 testing and HIV-1 viral load (VL) monitoring are needed for people living with HIV (PLHIV) in China. The Xpert HIV-1 VL assay was prequalified by the World Health Organization in 2017 but has not been evaluated in China. A multicenter evaluation was conducted to assess the accuracy of the Cepheid Xpert HIV-1 VL assay compared to the Abbott RealTime HIV-1 assay in China. Overall agreement was seen in 558 of 562 specimens (99.29%) with a κ value of 0.962. Pearson's coefficient between the two assays was 0.943. Analyzed by the Bland-Altman method, the mean bias was -0.54 log10 copies/mL, and 94.05% results fell within the 95% confidence limit of agreement (-1.248 to 0.168 log10 copies/mL). The coefficient of variation of the Cepheid Xpert HIV-1 VL assay ranged from 0.61% to 1.55%, as determined by testing eight positive plasma specimens with three different lots on different days. Due to its simplicity, random-access, rapid turnaround time, and accuracy, the Xpert HIV-1 VL assay can be used in local hospitals and clinics that bear the burden of identifying and treating HIV patients in China.