Response of Home-Use Adaptive Pressure Modes to Simulated Transient Hypoventilation
Frédéric Lofaso1,2, Karl Leroux3, Ghilas Boussaid2
1Department of Clinical Physiology, AP-HP Hôpital Raymond Poincaré, Université de Versailles Saint Quentin en Yvelines, Garches, France. f.lofaso@aphp.fr.
Background:
Adaptive servoventilation (ASV) is a recently developed ventilation mode designed to stabilize ventilation in patients with central sleep apnea and Cheyne-Stokes respiration. Alternatively, modes aiming to maintain average ventilation over several breaths, such as average volume-assured pressure support (AVAPS) and intelligent volume-assured pressure support (iVAPS), could be efficient during ventilation instability by reducing central events. These modes are available on a variety of devices. This bench evaluation studied the response of these different modes and devices to simulated transient hypoventilation events.
Methods:
Three home ventilation devices operating in ASV modes (AirCurve S10 VAuto, ResMed; DreamStation autoSV, Philips; Prisma CR, Löwenstein) and 2 ventilators with the AVAPS mode (DreamStation BiPAP, Philips; Lumis iVAPS, ResMed) were evaluated during transient central hypopnea/hypoventilation simulations characterized by a constant breathing frequency of 15 breaths/min and a progressive decrease of tidal volume (VT) from 500 mL to 50 mL, in 18, 12, 9, and 6 breaths, respectively, followed by a progressive return to the baseline at the same rate.
Results:
The AirCurve S10 VAuto reacted to a VT decrease between 80% and 50% of baseline VT. DreamStation BiPAP and Prisma CR reacted when VT decreased to between 60% and 30% of baseline VT, whereas the AVAPS response to hypopnea occurred during the crescendo phase of hypopnea/hypoventilation VT. The iVAPS response was between that of the AirCurve S10 VAuto and the other ASV devices. Among the ASV devices, the minimum VT was higher with AirCurve S10 VAuto, followed by the Prisma CR and the DreamStation BiPAP. Minimum VT was not influenced by AVAPS and was improved by iVAPS without outperforming the AirCurve S10 VAuto. Maximum VT was increased by iVAPS, whereas ASV devices did not induce a significant VT overshoot.
Conclusions:
ASV devices improved central hypopnea/hypoventilation events without inducing hyperpnea events and therefore were better adapted than AVAPS and iVAPS devices, with notable differences in their responses to hypoventilation events.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Hyperpnea and Hyperventilation
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...