Effect of X-ray beam energy and image reconstruction technique on computed tomography numbers of various tissue
1Department of Radiologic Sciences, Kuwait University, Sulaibikhat 90805, Kuwait.
Introduction:
Computed tomography (CT) numbers are used in radiological diagnosis, attenuation correction and radiotherapy treatment planning. Modern CT scanners use iterative reconstruction methods instead of the traditional filtered back projection (FBP). Hence, the investigation of CT number accuracy with image reconstruction techniques and X-ray tube potential (kVp) used in CT is warranted. The aim of this study is to evaluate the effect of Sinogram Affirmed Iterative Reconstruction (SAFIRE) Technique and image acquisition at different tube potentials on CT numbers of different tissue equivalent materials.
Methods:
Images of the Computerised Imaging Reference System Model 062M Electron Density Phantom were acquired at different tube potentials and reconstructed using FBP and different strengths of SAFIRE. Average CT numbers, in circular regions of interest, and their standard deviations were used to investigate any dependence of CT numbers on tube potentials and/or image reconstruction technique using non-parametric statistical tests with p-values set at 0.05.
Results:
Statistically significant differences in CT numbers were not observed (p > 0.091) between the different image reconstruction techniques. CT number of bone equivalent materials increased significantly (p < 0.015), by up to 400 Hounsfield Units, when tube potential was decreased. Such extent of CT number change over the tube potentials range used in this study may influence diagnostic outcomes in lung nodule, contrast enhanced and calcium score studies. For all other tissue equivalent materials, the CT number did not change significantly for different tube potentials. Linear relationship was observed between CT numbers and electron densities.
Conclusion:
The study concludes that the CT numbers of all tissues did not change significantly with image reconstruction methods. However, the CT numbers of bone equivalent materials increased with decreasing tube potentials, which may result in misrepresentation of clinical information obtained.
Implications For Practice:
When CT images are used to extract quantitative parameters such as calcium score, to characterise lung nodules and contrast enhanced structures, the kVp used for image acquisition should be carefully selected to avoid any misrepresentation of clinical information.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT