Clinical implications of elective replacement indicator setting changes in patients with dual-chamber pacemaker
Kari A Phillips1, Shiva Ponamgi2,3, Benjamin Mundell4
1Department of Pediatric and Adolescent Medicine, Mayo Clinic, Rochester, Minnesota, USA.
Pacemaker setting changes at elective replacement indicator (ERI) significantly increase patient symptoms and healthcare use. These changes include switching to single-chamber pacing or losing rate response (RR).
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Dual-chamber pacemakers may revert to single-chamber pacing or lose rate response (RR) at elective replacement indicator (ERI) to conserve battery.
- These modifications can potentially lead to adverse outcomes such as heart failure, atrial fibrillation (AF), and pacemaker syndrome.
Purpose of the Study:
- To investigate whether single-chamber pacing or loss of RR during ERI in dual-chamber pacemaker patients increases symptom burden, healthcare utilization, and atrial fibrillation (AF).
Main Methods:
- A retrospective analysis of 700 patients with dual-chamber pacemakers was conducted.
- Three comparisons were performed: mode change with RR loss vs. no change; RR loss only vs. no change; and mode change only vs. no change.
Main Results:
- Patients experiencing setting changes (mode change and/or RR loss) at ERI reported significantly more symptoms (dyspnea, fatigue) compared to those without changes (46% vs. 4%, p < .0001).
- Patients with setting changes also had increased provider contact (p = .0001), and all hospitalizations (2%) occurred in patients with setting changes.
- While mode change and RR loss together significantly increased symptoms and healthcare use, isolated RR loss or mode change showed less pronounced effects in some comparisons.
Conclusions:
- Elective replacement indicator (ERI) setting adjustments, including transitions to single-chamber pacing and/or loss of rate response (RR), are associated with a significant rise in patient-reported symptoms.
- These pacemaker parameter modifications also correlate with increased healthcare utilization, highlighting potential patient management challenges.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiac Catheterization IV: Nursing Management
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Endocarditis III: Medical Management