2D range modulator for high-precision water calorimetry in scanned carbon-ion beams
Kim Marina Holm1, Ulrich Weber, Yuri Simeonov
1Department of Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, D-38116 Braunschweig, Germany. Department of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, D-69120 Heidelberg, Germany. Heidelberg Institute for Radiation Oncology (HIRO), Heidelberg, Germany. Department of Physics and Astronomy, University of Heidelberg, Im Neuenheimer Feld 226, D-69120 Heidelberg, Germany.
Abstract:
Ionization chamber-based dosimetry for carbon-ion beams still shows a significantly higher standard uncertainty than high-energy photon dosimetry. This is mainly caused by the high standard uncertainty of the correction factor for beam quality [Formula: see text]. Due to a lack of experimental data, the given values for [Formula: see text] are based on theoretical calculations. To reduce this standard uncertainty, [Formula: see text] factors for different irradiation conditions and ionization chambers (ICs) can be determined experimentally by means of water calorimetry. To perform such measurements in a spread-out Bragg peak (SOBP) for a scanned carbon-ion beam, we describe the process of creating an almost cubic dose distribution of about 6 × 6 × 6 cm3 using a 2D range modulator. The aim is to achieve a field homogeneity with a standard deviation of measured dose values in the middle of the SOBP (over a lateral range and a depth of about 4 cm) below 2% within a scanning time of under 100 s, applying a dose larger than 1 Gy. This paper describes the optimization and characterization of the dose distribution in detail.
Related Concept Videos
Constant Volume Calorimetry
Constant Pressure Calorimetry