CT-based patient-specific instrumentation for total knee arthroplasty in over 700 cases: single-use instruments are
Stefan Gaukel1, Raphael N Vuille-Dit-Bille2, Michel Schläppi3
1Department of Orthopedics and Traumatology, Cantonal Hospital Winterthur, Brauerstrasse 15, 8409, Winterthur, Switzerland. stefan.gaukel@ksw.ch.
Purpose:
Efforts in total knee arthroplasty are made to improve accuracy for a correct leg axis and reduce component malpositioning using patient-specific instruments. It was hypothesized that use of patient-specific instruments (vs. computer-navigated and conventional techniques) will reduce the number of outliers. Our second hypothesis was that single-use instrumentation will lead to the same accuracy compared to patient-specific instruments made of metal.
Methods:
708 primary total knee arthroplasties between 2014 and 2018 using computer tomography (CT)-based patient-specific cutting block technique and a preoperative planning protocol were retrospectively reviewed. Preoperative data [hip-knee-angle (HKA), lateral distal femoral angle (LDFA), medial proximal tibial angle (MPTA), tibial slope, femoral component flexion] was compared to postoperative performed standard radiological follow-up X-rays. Differences of > 3° between measurements were defined as outliers.
Results:
Overall 500 prostheses using standard instrumentation and 208 prostheses using single-use instruments were implanted. Preoperative HKA axes (- 1.2°; p < 0.001), femoral component flexion (Δ 0.8°, p < 0.001), LDFA (Δ - 1.5°, p < 0.001), MPTA (Δ - 0.5°, p < 0.001) and tibial posterior slopes (Δ 0.5°, p < 0.001), respectively, were different from postoperative axes. More outliers occurred using standard (vs. single-use) instruments (p < 0.001) regarding postoperative HKA (ranges of standard- vs. single-use: instruments: HKA 178.0°-180.5° vs. 178.0°-180.5°, femoral component flexion 0.0°-6.0° vs. 0.0°-4.5°, LDFA 90.0°-91.0° vs. 90.0°-90.0°, MPTA 90.0°-90.0° vs. 90.0°-90.0°, tibial posterior slope - 10° to 10° vs. - 1° to 10°). No differences were seen for other angles measured. Comparing both systems, total number of outliers was higher using standard (8%) vs. single-use instruments (4.3%).
Conclusion:
This study shows a high accuracy of CT-based patient-specific instrumentation concerning postoperative achieved knee angles and mechanical leg axes. Single-use instruments showed a similar accuracy.
Level Of Evidence:
III.