High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Activity regimes for 82Rb cardiac PET: Effects on absolute MBF and MPI
Camilla M Hoff1, Jens Sørensen2,3, Nana L Christensen1
1Department of Nuclear Medicine & PET, Aarhus University Hospital, Aarhus, Denmark.
Background:
Selection of optimal tracer activity for 82Rb PET is based on a trade-off between necessary count-statistics in the late static phase and detector saturation in the early blood-pool phase. Administered tracer activity recommended in prescribing information differs substantially from recommendations in current literature. The present study examines the effect on both absolute myocardial blood flow (MBF), myocardial flow reserve (MFR) and relative myocardial perfusion imaging (MPI) of reducing dose.
Methods:
Forty patients were scanned twice on a PMT-based PET/CT (GE D690): At recommended activity (1110 MBq) and at either 740 or 370 MBq. MBF, MFR, total perfusion deficit (TPD) and ejection fractions (EF) were quantified. Results were compared using linear regression and Bland-Altman plots.
Results:
Linear correlation between MBF at 1110 MBq at either reduced activity had an R2 > 0.98. A small bias (± 5%-9%) was observed with opposite signs for 1110/740 and 1110/370. Limits of agreement for MBF were larger for 1110/370. MFR had a lower linear correlation (R2 = 0.96), but wide limits of agreement especially for 1110/370. TPD and EF correlated well at 1110/740 (R2 = 0.96 and 0.99, respectively), but large scatter was observed for 1110/370.
Conclusion:
Reduction of the tracer activity to 740 MBq, significantly reduced dead-time correction factors, while still producing reliable static and gated images. However, despite large dead-time at 1110 MBq, no systematic bias on absolute MBF was observed compared to reduced activities.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...