[Noninvasive hemoglobin monitoring in clinical trials: a systematic review and meta-analysis]
Tailyne Zortéa1, Daniele Paola da Silva Wizbicki1, Kristian Madeira2
1Universidade do Extremo Sul Catarinense, Criciúma, SC, Brasil.
Background And Objectives:
The measurement of Hb by co-oximetry is an innovative technique that offers efficiency and agility in the processing of information regarding the measurement of Hemoglobin concentration (Hb) obtained through continuous, non-invasive and rapid monitoring. Because of this attribute, it avoids unnecessary exposures of the patient to invasive procedures by allowing a reduction in the number of blood samples for evaluation and other unnecessary therapies. It also helps to make decisions about the need for transfusion and how to handle it. The objective of this study is to compare the performance offered to obtain Hb values between the Masimo Corporation (Irvine, CA, USA) instrument and the standard gold tool (laboratory examination).
Contents:
The study corresponds to a systematic review followed by meta-analysis, which included fully registered full-text clinical trials published from 1990 to 2018. PubMed, Cochrane, Medline, Embase and Web of Science databases were investigated. The mean overall difference found between the non-invasive and invasive methods of hemoglobin monitoring was 0.23 (95% CI -0.16, 0.62), that is, it did not present statistical significance (p = 0.250). The results of the analysis of heterogeneity within and between the studies indicated high levels of inconsistency (Q = 461.63, p < 0.0001, I2 = 98%), method for Hb values.
Conclusions:
Although the mean difference between noninvasive measurements of Hb and the gold standard method is small, the co-oximeter can be used as a non-invasive "trend" monitor in detecting unexpected responses at Hb levels.
More Related Videos
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy