Patient-Specific Computer Simulation in TAVR With the Self-Expanding Evolut R Valve
Nahid El Faquir1, Ole De Backer2, Johan Bosmans3
1Department of Cardiology, Thoraxcenter, Erasmus Medical Center, Rotterdam, the Netherlands.
Patient-specific computer simulation did not change valve size selection for transcatheter aortic valve implantation (TAVI). However, it influenced target depth decisions and procedural execution, aiding in achieving optimal outcomes.
Area of Science:
- Cardiovascular Medicine
- Medical Imaging
- Surgical Technology
Background:
- Transcatheter aortic valve replacement (TAVR) is increasingly used in lower-risk patients, necessitating optimal outcomes.
- Patient-specific computer simulation can predict TAVR outcomes, but its clinical utility and accuracy require prospective validation.
Purpose of the Study:
- To evaluate the added value and predictive accuracy of TAVIguide software in clinical TAVR practice.
- To assess the impact of patient-specific computer simulation on procedural decisions and outcomes.
Main Methods:
- A prospective, observational, multicenter study involving 80 patients with severe aortic stenosis undergoing TAVR with the Evolut R valve.
- Simulation was used in 42 patients to determine valve size and implantation depth, compared to standard methods in 38 patients.
- Primary endpoints included differences in valve size and implantation depth decisions; predictive power was assessed by comparing simulated and observed aortic regurgitation.
Main Results:
- Simulation did not alter valve size selection in 39/42 patients but affected implantation depth in 7/42.
- The TAVIguide simulation influenced the TAVR procedure in 16/42 patients, impacting decisions on additional measures.
- A trend towards higher predicted than observed aortic regurgitation was noted (17.5 vs. 12 ml/s, p=0.13).
Conclusions:
- Patient-specific computer simulation, like TAVIguide, did not significantly change valve size selection for TAVR.
- Simulation impacted the selection of target implantation depth and influenced procedural execution to achieve desired depth.
- The study suggests simulation can refine TAVR planning and execution, potentially improving outcomes.