Centering Ability of a New Nickel-Titanium Rotary Instruments with a Peculiar Flat-side Design: An In Vitro Study
Dario Di Nardo1, Gabriele Miccoli1, Alessandro Mazzoni1
1Department of Oral and Maxillofacial Sciences, Sapienza University of Rome, Rome, Italy.
Aim:
The aim of the present study was to investigate the role of the flat design compared with a non-flat designed instrument on the centering ability in a simulated root canal system.
Materials And Methods:
Two file systems were chosen: the F-One (Fanta Dental, Shanghai, China) and a prototype with the same design but without the flat-side design. A total of 50 simulated L-shaped root canals in resin blocks were shaped with a new instrument each. Pre- and postcanal preparation images were captured and superimposed to measure and compare the canal transportation and evaluate the centering ability of each file. The t test was performed and the level of significance was set at p < 0.05.
Results:
Prototype files removed significantly less resin from the inner and the outer parts of the canal and showed significantly higher mean transportation values.
Conclusion:
It can be concluded that the F-One file system demonstrated better shaping ability compared with the prototype file system in L-shaped resin-simulated canals.
Clinical Relevance:
The present study showed the safety of the flat-side design, minimizing the risk of iatrogenic errors such as ledging, perforation, or zipping of the root canal.