Related Experiment Video
Updated: Dec 14, 2025

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Comparison of dosimetric parameters of small-field electron beams between Advanced Markus, Semiflex 3D, and Diode E
Mohammad Reza Bayatiani1, Akbar Aliasgharzadeh2, Fatemeh Seif1
1Department of Medical Physics and Radiotherapy, Arak University of Medical Sciences and Khansari Hospital, Arāk, Iran.
Abstract:
The usage of dosimetry of small fields in radiotherapy to measure radiation dose is difficult because of high-dose gradients, lateral electronic disequilibrium, and detector volume effects. In this study, three dosimeters namely, Markus, Semiflex 3D, and Diode E were tested using the Elekta-accelerator electron beams. The electron beam parameters, penumbra, and output factor were determined using these dosimeters for each field size and energy. According to the results, Diode E and Advanced Markus exhibited the greatest difference in Rq among the electron beam parameters. Furthermore, the greatest difference in penumbra was observed between Diode E and Advanced Markus for the field size of 3 cm2 at 10 MeV. In terms of output factor, three dosimeters exhibited the greatest difference between Diode E and Advanced Markus for the field size of 3 cm2 at 10 MeV. The findings indicate that the Semiflex 3D can be regarded as an appropriate dosimeter for electron small-field dosimetry.

