Related Experiment Video
Updated: Dec 14, 2025

Murine Echocardiography of Left Atrium, Aorta, and Pulmonary Artery
Published on: February 20, 2017
Two- or 3-Dimensional Echocardiography-Derived Cardiac Output Cannot Replace the Pulmonary Artery Catheter in Cardiac
Lisa Q Rong1, Mudit Kaushal1, Elizabeth Mauer2
1Department of Anesthesiology, Weill Cornell Medicine/New York Presbyterian, New York, NY.
Objectives:
Three-dimensional (3D) transesophageal echocardiography (TEE) has been shown to be more accurate than 2D TEE for the evaluation of the left ventricular outflow tract area. The aim of the present study was to compare the agreement of 3D echocardiography-derived cardiac output (CO) with thermodilution-derived CO (TDCO) before and after cardiopulmonary bypass (CPB).
Design:
This was a prospective observational study of patients who underwent cardiac surgery between 2016 and 2018.
Setting:
Weill Cornell Medicine, a single large academic medical center.
Participants:
The study comprised 78 patients undergoing elective cardiac surgery.
Interventions:
CPB, TEE, pulmonary artery catheter, and elective cardiac surgery.
Measurements And Main Results:
Two-dimensional CO, 3D CO-diameter, and 3D CO-area values pre-CPB were strongly correlated with one another both pre-CPB and post-CPB. The 3D CO-diameter and the 3D CO-area were mildly correlated, with TDCO measurements pre-CPB (r = 0.46 and 0.39, respectively) and post-CBP (r = 0.43 and 0.47, respectively). Pre-CPB 3D CO-diameter had the most agreement with TDCO in terms of bias (-0.13 L/min); however, the limits of agreement (LOA) were wide (-2.2- to- 2.45 L/min). Post-CPB, 3D CO-diameter had the most agreement with TDCO in terms of bias (0.41) but with wide LOA (-3.29 to 2.47). All pre-CPB echocardiography-derived CO (2D CO, 3D CO-diameter, 3D CO-area) had more agreement with TDCO than did post-CPB measurements.
Conclusions:
Three-dimensional CO measurements were only modestly correlated with pulmonary artery catheter-derived CO pre-bypass and post-bypass. Despite low bias, the wide LOA from 2D CO, 3D CO-diameter, and 3D-area compared with TDCO suggested that the 2 methods are not interchangeable.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Acute Coronary Syndrome III: Diagnostic Studies
Cardiac Catheterization II: Right Heart Catheterization