Mean arterial pressure estimated by brachial pulse wave analysis and comparison with currently used algorithms
Andrea Grillo1, Paolo Salvi2, Giulia Furlanis1
1Azienda Sanitaria Universitaria Integrata of Trieste, Department of Medicine, Medicina Clinica Unit, Trieste.
Objective:
Mean arterial pressure (MAP) is usually calculated by adding one-third of pulse pressure (PP) to DBP. This formula assumes that the average value of pulse waveform is constant in all individuals and coincides with 33.3% of PP amplitude (MAP = DBP + PP × 0.333). Other formulas were lately proposed to improve the MAP estimation, adding to DBP an established percentage of PP: MAP = DBP + PP × 0.40; MAP = DBP + PP × 0.412; MAP = DBP + PP × 0.333 + 5 mmHg.
Methods:
The current study evaluated the integral of brachial pulse waveform recorded by applanation tonometry in 1526 patients belonging to three distinct cohorts: normotensive or hypertensive elderly, hypertensive adults, and normotensive adults.
Results:
The percentage of PP to be added to DBP to obtain MAP was extremely variable among individuals, ranging from 23 to 58% (mean: 42.2 ± 5.5%), higher in women (42.9 ± 5.6%) than men (41.2 ± 5.1%, P < 0.001), lower in the elderly cohort (40.9 ± 5.3%) than in the general population cohort (42.8 ± 6.0%, P < 0.001) and in the hypertensive patients (42.4 ± 4.8%, P < 0.001). This percentage was significantly associated with DBP (β = 0.357, P < 0.001) and sex (β = 0.203, P < 0.001) and significantly increased after mental stress test in 19 healthy volunteers (from 39.9 ± 3.2 at baseline, to 43.0 ± 4.0, P < 0.0001). The average difference between MAP values estimated by formulas, compared with MAP assessed on the brachial tonometric curve, was (mean ± 1.96 × SD): -5.0 ± 6.7 mmHg when MAP = DBP + PP × 0333; -1.2 ± 6.1 mmHg when MAP = DBP + PP × 0.40; -0.6 ± 6.1 mmHg when MAP = DBP + PP × 0.412; -0.4 ± 6.7 mmHg when MAP = DBP + PP × 0.333 + 5.
Conclusion:
Due to high interindividual and intraindividual variability of pulse waveform, the estimation of MAP based on fixed formulas derived from SBP and DBP is unreliable. Conversely, a more accurate estimation of MAP should be based on the pulse waveform analysis.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...