Weiguo Li1, Lianying Fang1, Jieqing Li1
1Institute of Radiation Medicine, Shandong Academy of Medical Sciences, Jinan, China.
Objective:
The aim of this study was to measure occupational exposure doses of technologists who dispense and inject radiopharmaceuticals in 7 positron emission tomography/computed tomography (PET/CT) departments. This was done with the goal to help improving protective designs in PET departments and/or establishing national protection standards.
Method:
Common LiF thermoluminescence dosimeters (TLDs) were placed on the chest and necklace of the technologists to monitor whole-body and thyroid doses, respectively. Ring TLDs were also worn on both index fingers to measure individual hand doses. All TLDs were assembled and measured once every 3 months for a total of 12 months. Additionally, we measured and compared the dose of TLDs attached to both the inside and the outside of the technologist's lead coat.
Results:
Technologists received relatively high exposures, which accounted for 64% to 94% of the collective dose in their respective departments. Their thyroid doses ranged from 1.2 to 1.7 mSv/a; some technologists' hand doses exceeded 500 mSv/a. Use of a lead coat reduced the average dose by 8%.
Conclusion:
Technologists working in PET/CT departments were the main population exposed to radiation. This work underscores the need for enhanced protective measures for these workers to better reduce their exposure, particularly for their hands.
More Related Videos
09:36Safety Precautions and Operating Procedures in an ABSL-4 Laboratory: 4. Medical Imaging Procedures
Published on: October 3, 2016
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography