Flow balance optimization and fluid removal accuracy with the Quanta SC+ hemodialysis system
Clive Buckberry1, Nicholas Hoenich2, Paul Komenda3
1Engineering Department, Quanta Dialysis Technologies Ltd , Alcester, UK.
Background:
Fluid management is integral to hemodialysis, both to correct abnormalities in a patient's plasma composition and to maintain fluid balance. Consequently, accurate net fluid removal during treatment is a critical design element of hemodialysis machines. As dialyzers have evolved, with increased ranges of ultrafiltration coefficients available, it has become more challenging for dialysis machines to minimize errors in flow balance and net fluid removal.
Research Design And Methods:
This paper describes the design, evaluation and experimental performance of the flow balance and ultrafiltration module of the SC+ system to deliver clinically specified fluid removal with both passive and active control measures, in laboratory conditions designed to simulate a wide range of therapies.
Results:
The use of passive and active control allows the errors to be minimized across a wider dynamic range of conditions. For the SC+ system, the average flow balance error was 1 mL/hr with an SD of 19 mL/hr and with ultrafiltration it was 13 mL/hr and an SD of 20 ml/hr across all conditions.
Conclusions:
This paper demonstrates that the SC+ hemodialysis system, a small, simple and versatile CE marked device, operates within the limits required by international standards across a wide range of experimental conditions.
More Related Videos
10:19Three-Dimensionally Printed Microfluidic Cross-flow System for Ultrafiltration/Nanofiltration Membrane Performance Testing
Published on: February 13, 2016
12:11A Modified EPA Method 1623 that Uses Tangential Flow Hollow-fiber Ultrafiltration and Heat Dissociation Steps to Detect Waterborne Cryptosporidium and Giardia spp.
Published on: July 9, 2012
Related Concept Videos
Hemodialysis III: Nursing Management
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Hemodialysis II: Procedure and Complications
Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Hemodialysis I: Introduction