Mechanical Evaluation of 2.7- Versus 3.5-mm Plating Constructs for Midshaft Clavicle Fractures
Mohammad M Alzahrani1, Adam Cota, Khalid Alkhelaifi
1From the Department of Orthopaedics (Dr. Alzahrani), College of Medicine, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia , the Rocky Mountain Orthopaedic Associates (Dr. Cota), Grand Junction, CO, the Department of Orthopedic Surgery (Dr. Alkhelaifi), ASPETAR, Orthopaedic and Sports Medicine Hospital, Doha, Qatar, and the Division of Orthopaedic Surgery (Dr. Harvey), McGill University Health Center, Montreal, Quebec, Canada.
Objectives:
This study compares the mechanical performance of 2.7- and 3.5-mm plating constructs for the treatment of midshaft clavicle fractures.
Methods:
Twenty-four synthetic clavicles were randomly divided into four treatment groups-Synthes 2.7-mm cold-worked calcaneal reconstruction plate with 6 (CRP6) or 8 bicortical screws (CRP8); Synthes 3.5-mm LCP reconstruction plate (RP; and Synthes 3.5-mm LCP precontoured superior-anterior clavicle plate (PCRP). All clavicles were plated, a wedge-shaped inferior cortical defect was created, and testing was performed using a cantilever bending model to determine bending stiffness and yield point for each construct.
Results:
Bending stiffness for the 3.5-mm PCRP construct was markedly higher when compared with the other three constructs, whereas the 3.5-mm RP construct was markedly stiffer than both of the 2.7-mm CR constructs. The yield point for the 3.5-mm PCRP construct was greater than the other three constructs; however, the yield point for the 2.7-mm CRP with six screws and with eight screws was higher than the 3.5-mm RP construct. The amount of displacement required to reach the yield point was highest for the 2.7-mm CRP with six screws. and this was markedly higher than the values for the other three constructs.
Discussion:
The 3.5-mm plates demonstrated increased bending stiffness compared with the 2.7-mm plates. Despite the lower resistance to bending forces, the cold-worked 2.7-mm plate exhibited a markedly higher yield point and required markedly more superior to inferior displacement to initiate plastic deformation when compared with the 3.5-mm LCP RP.
Level Of Evidence:
Level IV.