Objective assessment of the combined effect of exomass-related- and motion artefacts in cone beam CT
Matheus Lima Oliveira1, Amanda P Candemil1, Deborah Q Freitas1
1Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas (FOP/Unicamp), Piracicaba-SP, Brazil.
Objectives:
To assess quantitatively the combined effect of exomass-related- and motion artefacts on voxel value parameters in cone beam CT (CBCT).
Methods:
A cylindrical phantom was manufactured, containing 21 tubes filled with a radiopaque solution, allowing the inclusion of three titanium implants in the periphery to induce exomass-related artefacts. The phantom was mounted on a robot simulating 0.75-, 1.5-, and 3 mm movements (nodding/lateral rotation/tremor). CBCT images with/without exomass and with/without movements were acquired in duplicate in three units: Cranex 3Dx, Orthophos SL-3D, and X1 (with motion-artefact correction). A cylindrical volume of interest was defined in each tube and voxel value mean and standard deviation were assessed. For each CBCT volume, the 21 mean voxel values were averaged providing the overall mean voxel value (MVV), and the standard deviation (among the 21 values) was calculated providing overall voxel value inhomogeneity (VVI). The standard deviation from each of the 21 volumes-of-interest were averaged, providing overall image noise (IN). The effect of the diverse tested situations was inferred from a repeated-measures analysis of variance, followed by Sidak's test (α = 0.05).
Results:
Overall, images acquired with exomass showed significantly (p ≤ 0.05) lower MVV, and higher VVI and IN. Motion artefacts aggravated exomass-related alterations. MVV and VVI were mostly affected by 3 mm nodding movements. Motion-artefact correction eliminated the deleterious effect of movement.
Conclusions:
CBCT voxel-value parameters are altered by exomass-related artefacts, and this finding is aggravated in the presence of motion artefacts. Motion-artefact correction effectively eliminated the deleterious impact of movement.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging