Related Experiment Video
Updated: Dec 13, 2025

Three-Dimensional Phase Resolved Functional Lung Magnetic Resonance Imaging
Published on: June 21, 2024
Enhanced POCS reconstruction for partial Fourier imaging in multi-echo and time-series acquisitions
Peter J Koopmans1,2, Viktor Pfaffenrot1,2
1Erwin L. Hahn Institute for Magnetic Resonance Imaging, University of Duisburg-Essen, Essen, Germany.
Purpose:
To improve partial Fourier (PF) imaging reconstruction in time-series or multi-echo acquisitions.
Methods:
Many PF methods use a phase estimate to restore Hermitian symmetry before filling missing k-space entries with measured data from the opposite half. This estimate is obtained from the symmetrically sampled, central part of k-space and its low-resolution results in artifacts near high-frequency phase effects (eg, tissue boundaries, vessels), limiting PF undersampling. Enhanced projection onto convex sets (POCS) uses full-resolution phase estimates and relies on alternating the half of k-space that is acquired in time series or multi-echo acquisitions. This enables full-resolution phase estimates to be calculated for each volume/echo, which are fed into the POCS framework. We apply enhanced POCS to high-resolution multi-echo FLASH and 3D-EPI functional MRI time-series data.
Results:
Reconstruction errors and their bias dramatically reduce compared with existing methods, without leading to temporal blurring in time-series acquisitions. This allows for higher PF acceleration factors at virtually no cost.
Conclusion:
Enhanced POCS results in superior PF reconstructions. Furthermore, as the resolution of the phase estimate used for symmetry correction no longer depends on the PF factor, enhanced POCS is more robust against larger PF omission.
Related Concept Videos
Reconstruction of Signal using Interpolation
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Aliasing
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...