Forceps through a 14F flexible sheath for safe retrieval of a dislodged left atrial appendage occluder
Libin Qiu1,2, Bing Rong1, Kai Zhang1
1The Key Laboratory of Cardiovascular Remodeling and Function Research, Chinese Ministry of Education, Chinese National Health Commission, and Chinese Academy of Medical Sciences, The State and Shandong Province Joint Key Laboratory of Translational Cardiovascular Medicine, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, China.
Abstract:
The LAmbre occluder (Lifetech Scientific, Shenzhen, China), a new device for left atrial appendage occlusion (LAAO), is increasingly used, but the retrieval procedure for dislodgements has rarely been reported in humans. We herein reported a case of safe retrieval of dislodged LAmbre occluder. The occluder dislodged to the left atrium (LA) at the end of the LAAO procedure. We failed to retrieve the device in the LA using a LASSO catheter and forceps through an 8.5F sheath. After it suddenly flowed into the aortic arch, we successfully retrieved it using the same forceps through a 14F flexible sheath. We also discuss the reasons for device dislodgment and experiences of device retrieval. We think that forceps can be used to retrieve an embolized LAmbre device in the aorta through a 14F flexible sheath.