Metallosis after knee replacement: a review
Ismail Sahan1, Konstantinos Anagnostakos2
1Zentrum für Orthopädie und Unfallchirurgie, Klinikum Saarbrücken, Winterberg 1, 66119, Saarbrücken, Germany. i.sahan@outlook.de.
Introduction:
Although metallosis is a well-known complication after total hip arthroplasty, little is known about this phenomenon after total (TKA) or unicompartmental knee arthroplasty (UKA). The aim of the present work was to review the current knowledge about the reasons and the diagnostic as well as therapeutic management of metallosis after knee arthroplasty.
Materials And Methods:
A literature search was performed through PubMed until April 2019. Search terms were "metallosis" in combination with "knee", "knee prosthesis", "knee arthroplasty" and "knee replacement", respectively. All publications were analyzed regarding publication year, level of evidence, number of knees/patients treated, type of prosthesis, metallosis cause, time period between primary implantation and metallosis emergence, laboratory examination, treatment, complications and follow up.
Results:
A total of 38 studies reporting on a total of 97 knees were identified. 29 studies reported on metallosis after TKA, 8 after UKA, and one study after both procedures. The time period between the primary implantation and metallosis emergence ranged between 6 weeks and 26 years. The most common reason was the failure of a metal-backed patellar component in 40%, followed by implant/structural- and PE failure (wear/dislocation) in 27% and 18% of the cases, respectively. Complete blood cell count, serum chemistry, erythrocyte sedimentation rate or C-reactive protein serum values were not indicative to diagnose metallosis. The diagnosis was confirmed by histopathological analyses and macroscopic evaluation during surgery. Depending on the particular cause various surgical procedures have been performed. Complete prosthesis exchange was the most common one showing no complications in 89.4% of the cases.
Conclusions:
Metallosis after knee arthroplasty is a rare and perhaps underestimated or under published complication. A systematic diagnostic approach is necessary for the timely and correct diagnosis. A thorough debridement as well as a (sub)total synovectomy should be always performed. In cases with a damaged component, a partial/complete prosthesis exchange leads to the best results. Should a malalignment be the cause of the metallosis, then it should be corrected within the revision surgery.