Evaluation method and platform of vibrational disturbance test for ventricular assist devices
Shu Li1, Juan Yan2, Tingting Wu3
1China National Institutes for Food and Drug Control, Institute for Medical Device Control, Beijing, China.
Objective:
For ventricular assist devices utilizing levitation bearing technology, collision and wear of the rotor would occur if the bearing cannot resist disturbances from patient activities and vibration/shock impacts. The reliability of ventricular assist devices can be seriously affected, potentially impairing blood compatibility (e.g. thrombus generation) and threatening patient safety. In this article, we proposed a vibrational disturbance test protocol for ventricular assist devices.
Methods:
Two kinds of vibrational disturbances-translational and rotational-were defined and the disturbance levels determined as 6G and 10 rad/s per possible patient activity impact. A test platform was built according to the disturbance level requirements.
Results:
The test platform successfully generated the required disturbance. The vibration test evaluation criteria for ventricular assist device assistance (characterized by the pressure head, flow rate, and driving current waveform) were well verified.
Conclusion:
Compared with translational vibration, rotational vibration had a greater impact on the rotor. Accurate control of high-speed rotor is difficult because of the gyroscopic effect. As conventional random vibration tests cannot cover all risk situations, it is recommended that translational and rotational disturbance tests are carried out for levitation ventricular assist devices. We recommend that researchers and manufacturers pay attention to the vibrational impact of rotor-levitated ventricular assist devices.
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Mechanical Ventilation I: Indication and Settings
Mitral Stenosis II: Clinical features and Diagnostic Tests