Static and Dynamic Stress Analysis of Standard- and Narrow-Diameter Implants: A 3D Finite Element Analysis
Purpose:
The purpose of the three-dimensional (3D) finite element analysis study was to compare the use of titanium-zirconium (Ti-Zr) narrow-diameter implants as an alternative to titanium (Ti) or Ti-Zr standard-diameter implants in the posterior jaw regions.
Materials And Methods:
Ti-Zr and Ti standard-diameter implants (4.1 mm) and Ti-Zr narrow-diameter implants (3.3 mm) in cylindrical (parallel) macrodesign were simulated in the mandibular and maxillary first premolar area. Forces of 100 N were applied to the crowns in a vertical and oblique (45-degree angle to the long axis) direction. The von Mises stresses and fatigue strength values of the implants and principal stresses in the bone structures were evaluated.
Results:
In vertical and oblique force application, stress data in cortical and trabecular bone structures were found to be higher in 3.3-mm Ti-Zr narrow-diameter implant models than 4.1-mm Ti-Zr and Ti standard-diameter implant models. Also, the von Mises stress data of the 3.3-mm Ti-Zr narrow-diameter implants were higher than the 4.1-mm Ti-Zr and Ti standard-diameter implants. The shortest cycle of fatigue failure and estimated duration of clinical success (years) results were found in the mandibular 3.3-mm Ti-Zr implant model under oblique force, and these results remained below the identified 30-year critical threshold.
Conclusion:
Considering all implant models, Ti-Zr narrow-diameter implants exhibited higher stress values than Ti-Zr and Ti standard-diameter implants. In the premolar region, care should be taken biomechanically when using Ti-Zr narrow-diameter implants as an alternative to standard-diameter implants. Further comparative and in vivo studies are needed to examine the long-term success of Ti-Zr narrow-diameter implants as an alternative to standard-diameter implants.
More Related Videos
11:28A Coupled Experiment-finite Element Modeling Methodology for Assessing High Strain Rate Mechanical Response of Soft Biomaterials
Published on: May 18, 2015
07:16Author Spotlight: Development of a Novel Finite Element Analysis Model for Improved Orthognathic Surgical Techniques
Published on: October 20, 2023
Related Concept Videos
Three-Dimensional Analysis of Strain
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....
Stress Concentrations in Circular Shafts
Applications of Stress
The...
Normal Strain under Axial Loading