Related Experiment Video
Updated: Dec 13, 2025

Nanosensors to Detect Protease Activity In Vivo for Noninvasive Diagnostics
Published on: July 16, 2018
Evaluating the sensing performance of nanopore blockade sensors: A case study of prostate-specific antigen assay
Yanfang Wu1, Kyloon Chuah1, J Justin Gooding1
1School of Chemistry, Australian Centre for NanoMedicine, Australian Research Council Centre of Excellence in Convergent Bio-Nano Science and Technology, The University of New South Wales, Sydney, New South Wales, 2052, Australia.
Abstract:
The detection principle of nanopore sensors relies on measuring changes in electrical signal as analyte molecules translocate through a nanoscale pore. There are two challenges with this experimental construct when using nanopores for quantitative sensing with low detection limits in complex samples. The first is getting the analyte to the nanopore in a reasonable time frame and the second is other species in the sample also translocating through the nanopore and generating erroneous signals. We have developed a nanopore blockade sensor that alleviates the limitations of diffusion-limited mass transport and non-specific signals. Antibody-modified magnetic nanoparticles are utilized to deliver analytes of interest extracted from sample to an array of antibody-modified nanopores under a controlled electromagnet, resulting in long-term nanopore blocking events due to the formation of sandwiched immunocomplexes. Herein, this study reports on understanding some of important parameters in determining the performance of nanopore blockade sensing system, where prostate-specific antigen (PSA) is used as a model analyte. We describe the characterization of nanopore blockade sensing of PSA by (1) tuning on/off the electromagnet, (2) varying nanopore number in a nanopore chip, and (3) deploying the sensor in human plasma. Results show that magnetophoresis effectively facilitates active delivery and selective sensing of PSA to the nanopore. Nanopore chips with a larger number of nanopores are shown to receive more nanopore blockades for a given concentration of analyte. Furthermore, identifiable blockade events accounted for successful detection of PSA in plasma, indicate the high specificity of the sensing system.