Related Experiment Video
Updated: Dec 13, 2025

An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Compressed sensing for reduced hardware footprint in medical ultrasound
Jovan Mitrovic1, Zeljko Ignjatovic1, Lynn La Pietra2
1Dept of Electrical and Computer Engineering, University of Rochester, Rochester, NY, USA.
Abstract:
In this work, a compressed sensing method to reduce hardware complexity of ultrasound imaging systems is proposed and experimentally verified. We provide clinical evaluation of the method with a possible high compression rates (up to 64 RF signals compressed into a single channel on receive) which uses elastic net estimation for decoding stage. This allows a reduction in size and power consumption of the front-end electronics with only a minor loss in image quality. We demonstrate an 8-fold receive channel count reduction with a 3.16 dB and 3.64 dB mean absolute error for gallbladder and kidney images, respectively, as well as 7.4% increase in the contrast-to-noise ratio for kidney images and 0.1% loss in the contrast-to noise ratio for gallbladder images, on average. The proposed method may enable a fully portable ultrasonic device with virtually no loss in image quality as compared to a full size clinical scanner to be constructed.