Validation of Therapeutic Agent Conjugation to Polyvinyl Alcohol-Coated Medical Devices
Published on: November 29, 2024
Do not flick or drop off-label use plastic syringes in handling therapeutic proteins before administration
Nam Ah Kim1, Dong Jun Kim1, Seong Hoon Jeong1
1College of Pharmacy, Dongguk University-Seoul, Gyeonggi 10326, Republic of Korea.
Abstract:
This study investigates the formation of subvisible particles formed by external stresses produced such as flicking and dropping syringes. Flow imaging was used to visualize and quantify microparticles from 1 μm to over 25 μm as a result of mishandling. Microparticles increased in the presence of silicone oil that was present in syringes. Thus, silicone oil in syringes may affect the activity of therapeutic proteins being injected. Present data showed detailed and differentiated morphologies of proteinaceous particles, silicone oil, air bubbles, and plastic debris in mishandled syringes. In some cases, the presence of bisphenol A in syringes was detected by FT-IR. Disposable plastic syringes were evaluated and showed differences in their content of silicone oil. Syringes that contain 0.45 μm filters inside the needle cap as well as silicone oil-free syringes release proteinaceous subvisible particles after mechanical stress. These stress-generated particles can be delivered to patients, compromising patient care.
More Related Videos
08:06Automated Acoustic Dispensing for the Serial Dilution of Peptide Agonists in Potency Determination Assays
Published on: November 10, 2016
08:43Recombinant Collagen I Peptide Microcarriers for Cell Expansion and Their Potential Use As Cell Delivery System in a Bioreactor Model
Published on: February 7, 2018