Ventilator Failure Following Cardiopulmonary Bypass: A Case Report
Jerry J Markar1, Andrew A Disque
1From the Department of Anesthesiology and Perioperative Medicine, University of California, Los Angeles, Los Angeles, California.
Abstract:
We present 3 cardiothoracic surgery cases during which the Drager Perseus A500 anesthesia machine and the iNOmax Delivery System were used concurrently, resulting in ventilator failures while attempting to separate from cardiopulmonary bypass (CPB). The failure was found to be caused by the iNOmax Delivery System sampling line, where negative pressure on the Perseus A500 breathing circuit precipitated a ventilator failure when the machine was brought out of "standby mode." To our knowledge, there is no discussion of this interaction in the literature, and we find this to be an important issue for patient safety during cardiothoracic surgery.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-III
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...