Determination of machine-specific tolerances using statistical process control analysis of long-term uniform scanning
Suresh Rana1, Colton Eckert1, Hardev Singh1
1Department of Medical Physics, Oklahoma Proton Center, Oklahoma City, OK, USA.
Purpose:
The purpose of this study was twofold: (a) report the long-term monthly quality assurance (QA) dosimetry results of the uniform scanning beam delivery system, and (b) derive the machine-specific tolerances based on the statistic process control (SPC) methodology and compare them against the AAPM TG224 recommended tolerances.
Methods:
The Oklahoma Proton Center has four treatment rooms (TR1, TR2, TR3, and TR4) with a cyclotron and a universal nozzle. Monthly QA dosimetry results of four treatment rooms over a period of 6 yr (Feb 2014-Jan 2020) were retrieved from the QA database. The dosimetry parameters included dose output, range, flatness, and symmetry. The monthly QA results were analyzed using the SPC method, which included individuals and moving range (I-MR) chart. The upper control limit (UCL) and lower control limit (LCL) were set at 3σ above and below the mean value, respectively.
Results:
The mean difference in dose output was -0.3% (2σ = ±0.9% and 3σ = ±1.3%) in TR1, 0% (2σ = ±1.4% and 3σ = ±2.1%) in TR2, -0.2% (2σ = ±1.0% and 3σ = ±1.6%) in TR3, and -0.5% (2σ = ±0.9% and 3σ = ±1.3%) in TR4. The mean flatness and symmetry differences of all beams among the four treatment rooms were within ±1.0%. The 3σ for the flatness difference ranged from ±0.5% to ±1.2%. The 3σ for the symmetry difference ranged from ±0.4% to ±1.4%. The SPC analysis showed that the 3σ for range 10 cm (R10), R16, and R22 were within ±1 mm, whereas the 3σ for R28 exceeded ±1 mm in two rooms (3σ = ±1.9 mm in TR2 and 3σ = ±1.3 mm in TR3).
Conclusion:
The 3σ of the dose output, flatness, and symmetry differences in all four rooms were comparable to the TG224 tolerance (±2%). For the uniform scanning system, if the measured range is compared against the requested range, it may not always be possible to achieve the range difference within ±1 mm (TG224) for all the ranges.
More Related Videos
09:45Large-area Scanning Probe Nanolithography Facilitated by Automated Alignment and Its Application to Substrate Fabrication for Cell Culture Studies
Published on: June 12, 2018
05:04Author Spotlight: Introduction to Active Probe Atomic Force Microscopy with Quattro-Parallel Cantilever Arrays
Published on: June 13, 2023
Related Concept Videos
Introduction to Statistical Process Control
Quality Assurance
Interpreting X̄ Charts
An x̄ chart plots the values of individual measurements over time against control limits calculated from historical data. The central line...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...