Design and In Vivo Evaluation of a Non-Invasive Transabdominal Fetal Pulse Oximeter
Objective:
Current intrapartum fetal monitoring technology is unable to provide physicians with an objective metric of fetal well-being, leading to degraded patient outcomes and increased litigation costs. Fetal oxygen saturation (SpO2) is a more suitable measure of fetal distress, but the inaccessibility of the fetus prior to birth makes this impossible to capture through current means. In this paper, we present a fully non-invasive, transabdominal fetal oximetry (TFO) system that provides in utero measures of fetal SpO2.
Methods:
TFO is performed by placing a reflectance-mode optode on the maternal abdomen and sending photons into the body to investigate the underlying fetal tissue. The proposed TFO system design consists of a multi-detector optode, an embedded optode control system, and custom user-interface software. To evaluate the developed TFO system, we utilized an in utero hypoxic fetal lamb model and performed controlled desaturation experiments while capturing gold standard arterial blood gases (SaO2).
Results:
Various degrees of fetal hypoxia were induced with true SaO2 values ranging between 10.5% and 66%. The non-invasive TFO system was able to accurately measure these fetal SpO2 values, supported by a root mean-squared error of 6.37% and strong measures of agreement with the gold standard.
Conclusion:
The results support the efficacy of the presented TFO system to non-invasively measure a wide-range of fetal SpO2 values and identify critical levels of fetal hypoxia.
Significance:
TFO has the potential to improve fetal outcomes by providing obstetricians with a non-invasive measure of fetal oxygen saturation prior to delivery.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring pulse
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation