Related Experiment Video
Updated: Dec 13, 2025

Author Spotlight: Advancing Awake Nasotracheal Intubation with Flexible Video Rhino-Laryngoscopes
Published on: August 2, 2024
A new self-designed "tongue root holder" device to aid fiberoptic intubation
Xiaofei Cao1, Junbei Wu1, Yin Fang1
1Department of Anesthesiology, The First Affiliated Hospital with Nanjing Medical University, 300 Guangzhou Road, Nanjing, 210029, People's Republic of China.
Objective:
In this study, we aimed to assess the feasibility of fiberoptic intubation (FOI), using a new, self-designed, "tongue root holder" device, in combination with the jaw thrust maneuver.
Methods:
Three hundred patients undergoing elective surgery requiring orotracheal intubation were enrolled. Patients presented at least one or more risk factors for difficult airway. The patients were randomly allocated at a 1:1 ratio to one of two groups: group L, FOI with tongue root holder, or group C, standard FOI. Orotracheal FOI was performed after commencement of anesthesia. The jaw thrust maneuver was applied in both groups to facilitate advancement of the fiberoptic bronchoscope. The primary endpoint was the feasibility of FOI. The secondary endpoints were number of attempts, time to intubation, and airway clearance at the soft palate and epiglottis levels.
Results:
The FOI was achieved in all 150 patients in group L, significantly higher than that in group C (100% vs 95.3%; P = 0.015). Less attempts of intubation were made in group L (P = 0.039). Mean time to successful intubation on the first attempt was shorter in group L (P < 0.001). The mean times to view the vocal cord and carina were also shorter in group L (P = 0.011 and P < 0.001, respectively). Airway clearance was better in group L at both the soft palate and the glottis levels (P = 0.010 and P = 0.038, respectively).
Conclusions:
This study shows that FOI is feasible with the newly introduced, self-designed, "tongue root holder" device, when combined with the jaw thrust maneuver in patients with risk factors for difficult airway. The device also provides better airway clearance, less intubation attempts, and shorter time to intubation at first attempt.
Clinical Relevance:
Fiberoptic bronchoscope has been the gold standard for routine management of difficult airway. A technique to open the airway is introduced to reduce the incidence rate of upper airway obstruction.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Suctioning the Nasopharyngeal Airway
Equipment Required