Thermal impact of replacing constant voltage by low-frequency sine wave voltage in RF ablation computer modeling
Juan J Pérez1, Ana González-Suárez2, Enrique Nadal3
1BioMIT, Department of Electronic Engineering, Universitat Politècnica de València, Valencia, Spain.
Background And Objectives:
A constant voltage (DC voltage) is usually used in radiofrequency ablation (RFA) computer models to mimic the radiofrequency voltage. However, in some cases a low frequency sine wave voltage (AC voltage) may be used instead. Our objective was to assess the thermal impact of replacing DC voltage by low-frequency AC voltage in RFA computer modeling.
Methods:
A 2D model was used consisting of an ablation electrode placed perpendicular to the tissue fragment. The Finite Element method was used to solve a coupled electric-thermal problem. Quasi-static electrical approximation was implemented in two ways (both with equivalent electrical power): (1) by a constant voltage of 25 V in the ablation electrode (DC voltage), and (2) applying a sine waveform with peak amplitude of 25√2 V (AC voltage). The frequency of the sine signal (fAC) varied from 0.5 Hz to 50 Hz.
Results:
Sine wave thermal oscillations (at twice the fAC frequency) were observed in the case of AC voltage, in addition to the temperature obtained by DC voltage. The amplitude of the oscillations: (1) increased with temperature, remaining more or less constant after 30 s; (2) was of up to ±3 °C for very low fAC values (0.5 Hz); and (3) was reduced at higher fAC values and with distance from the electrode (almost negligible for distances > 5 mm). The evolution of maximum lesion depth and width were almost identical with both DC and AC.
Conclusions:
Although reducing fAC reduces the computation time, thermal oscillations appear at points near the electrode, which suggests that a minimum value of fAC should be used. Replacing DC voltage by low-frequency AC voltage does not appear to have an impact on the lesion depth.
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
07:17Non-equilibrium Microwave Plasma for Efficient High Temperature Chemistry
Published on: August 1, 2017
Related Concept Videos
Standing Waves in a Cavity
Sinusoidal Sources
In homes, the power supplies use sinusoidal sources to provide electricity. These sources generate a voltage that varies sinusoidally...
Source Transformation for AC Circuits