Rate-dependent pacing failure after pacemaker implantation: Novel insights into the mechanism of using adenosine
Hisaharu Ohe1, Yasushi Oginosawa1, Yasunobu Yamagishi1
1The Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyushu, Japan.
Abstract:
An 82-year-old woman received pacemaker implantation for sick sinus syndrome. Two days after the implantation, electrocardiography showed 2:1 atrial pacing failure, followed by a bradycardia-dependent increase in the atrial pacing threshold during a pacemaker examination. However, transient 1:1 atrial pacing capture recovered by adenosine triphosphate (ATP) administration, which was performed to evaluate the bradycardia-dependent pacing failure mechanism. We considered this phenomenon to be caused by Phase 4 depolarization and avoided replacing this atrial lead. Three weeks later, the atrial pacing threshold had improved. We report the potential role of Phase 4 depolarization in a bradycardia-dependent increase in pacing threshold by using ATP.
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
Electrophysiology of Normal Cardiac Rhythm
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...