Accuracy and Reliability of the Ankle Brachial Index Measurement Using a Multicuff Oscillometric Device Versus the
Shigeo Ichihashi1, Ileana Desormais2, Tomoko Hashimoto3
1Department of Radiology, Nara Medical University, Kashihara, Nara, Japan.
Objective:
Ankle brachial index (ABI) is widely used for the diagnosis of lower extremity artery disease (LEAD). The purpose of this prospective study was to validate the diagnostic ability and reproducibility of a four cuff automated oscillometric device vs. the Doppler method.
Methods:
Patients with suspected LEAD or asymptomatic individuals at risk because of the presence two or more cardiovascular risk factors were enrolled. For each patient, Doppler and oscillometric ABI measurements were repeated by two observers to address intra- and interobserver reproducibility.
Results:
In total, 118 patients were evaluated. The prevalence of Doppler ABI (Dop-ABI) ≤ 0.90 was 45.8%. Taking the Dop-ABI as the reference, the sensitivity, specificity, accuracy, positive and negative predictive values of oscillometric ABI (Osc-ABI) during the first measurement by the first observer were 89.1%, 94.4%, 94.1%, 91.8%, and 92.4%, respectively. The concordance for diagnosing ABI ≤0.90 between methods was excellent (kappa coefficients ranging from 0.80 to 0.88 with different observers). Intra-observer reproducibility assessed by intraclass correlation coefficient (ICC) between methods were 0.94 for observer 1 and 0.96 for observer 2. The intra-observer reproducibility using the same method was also excellent (ICC 0.94, 95% confidence interval [CI] 0.91-0.95) for Dop-ABI and 0.95 (95% CI 0.93-0.97) for Osc-ABI). The ICC for interobserver reproducibility using the same method was 0.95 (95% CI 0.92-0.96) for Dop-ABI and 0.96 (95% CI 0.94-0.97) for Osc-ABI.
Conclusion:
This study validates the excellent diagnostic performances of a four cuff oscillometric device specifically designed for screening for LEAD. The simple measurement method could therefore be advocated in primary care where fast, easy, and reliable methods are suitable.
More Related Videos
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation