His bundle pacing-is it the final frontier of physiological pacing ?-A single centre experience from the Indian
V S Prakash1, Anupama V Hegde1, U M Nagamalesh1
1Dept of Cardiology, M S Ramaiah Medical College, Bangalore, India.
Background:
Long term right ventricular pacing can have deleterious effects on left ventricular (LV) function. His bundle pacing (HBP), a novel procedure can probably circumvent this setback. We investigated if (1) HBP is associated with pacing induced LV dysfunction by using LV global longitudinal strain (GLS) and (2) intermediate term performance of the Select Secure (3830) lead in the His bundle location. This report is probably the first on HBP in the Indian population.
Methods:
61 patients, with normal LV ejection fraction (EF) with a guideline based indication for permanent pacing underwent a HBP pacemaker implantation using the His Select Secure 3830 lead; with lead guided mapping for locating the His bundle. The patients underwent GLS assessment; evaluation of the His lead parameters - sensing, impedance and capture thresholds immediately after implantation and at 6 months in addition to the standard follow up.
Results:
At 6 month follow up, the average GLS did not show significant variation from baseline in patients requiring ventricular pacing more than 40% and was similar, irrespective of selective or non selective His bundle pacing. All the patients had stable pacemaker parameters - with little change in capture threshold, lead impedance or sensing of the His bundle lead - implying electrical and mechanical stability on intermediate term follow-up.
Conclusion:
HBP is a feasible procedure in the hands of an experienced operator, with stable lead performance. It does not appear to be associated with pacing mediated left ventricular dysfunction at intermediate term follow up. It should probably become the default method of permanent pacing.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Special considerations while measuring pulse
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Pulse Assessment Sites