Myelin Imaging in Human Brain Using a Short Repetition Time Adiabatic Inversion Recovery Prepared Ultrashort Echo
Ya-Jun Ma1, Hyungseok Jang1, Zhao Wei1
1From the Departments of Radiology (Y.J.M., H.J., Z.W., Z.C., Y.X., R.R.L., E.Y.C., G.M.B., J.D.) and Neurosciences (J.C.B.) University of California San Diego, 9452 Medical Center Dr, La Jolla, CA 92037; and Radiology Service, Veterans Affairs San Diego Healthcare System, San Diego, Calif (E.Y.C.).
Abstract:
Background Water signal contamination is a major challenge for direct ultrashort echo time (UTE) imaging of myelin in vivo because water contributes most of the signals detected in white matter. Purpose To validate a new short repetition time (TR) adiabatic inversion recovery (STAIR) prepared UTE (STAIR-UTE) sequence designed to suppress water signals and to allow imaging of ultrashort T2 protons of myelin in white matter using a clinical 3-T scanner. Materials and Methods In this prospective study, an optimization framework was used to obtain the optimal inversion time for nulling water signals using STAIR-UTE imaging at different TRs. Numeric simulation and phantom studies were performed. Healthy volunteers and participants with multiple sclerosis (MS) underwent MRI between November 2018 and October 2019 to compare STAIR-UTE and a clinical T2-weighted fluid-attenuated inversion recovery sequence for assessment of MS lesions. UTE measures of myelin were also performed to allow comparison of signals in lesions and with those in normal-appearing white matter (NAWM) in patients with MS and in normal white matter (NWM) in healthy volunteers. Results Simulation and phantom studies both suggest that the proposed STAIR-UTE technique can effectively suppress long T2 tissues with a broad range of T1s. Ten healthy volunteers (mean age, 33 years ± 8 [standard deviation]; six women) and 10 patients with MS (mean age, 51 years ± 16; seven women) were evaluated. The three-dimensional STAIR-UTE sequence effectively suppressed water components in white matter and selectively imaged myelin, which had a measured T2* value of 0.21 msec ± 0.04 in the volunteer study. A much lower mean UTE measure of myelin proton density was found in MS lesions (3.8 mol/L ± 1.5), and a slightly lower mean UTE measure was found in NAWM (7.2 mol/L ± 0.8) compared with that in NWM (8.0 mol/L ± 0.8) in the healthy volunteers (P < .001 for both comparisons). Conclusion The short repetition time adiabatic inversion recovery-prepared ultrashort echo time sequence provided efficient water signal suppression for volumetric imaging of myelin in the brain and showed excellent myelin signal contrast as well as marked ultrashort echo time signal reduction in multiple sclerosis lesions and a smaller reduction in normal-appearing white matter compared with normal white matter in volunteers. © RSNA, 2020 Online supplemental material is available for this article. See also the editorial by Messina and Port in this issue.
Insights
A new STAIR-UTE MRI sequence effectively suppresses water signals to image myelin in the brain. This technique shows reduced myelin in multiple sclerosis lesions and normal-appearing white matter, aiding disease assessment.
Area of Science:
- Neuroimaging
- Biophysics
- Medical Physics
Background:
- Direct ultrashort echo time (UTE) imaging of myelin is challenged by water signal contamination in white matter.
- Water constitutes the majority of signals detected in white matter, hindering myelin visualization.
Purpose of the Study:
- To validate a novel short repetition time (TR) adiabatic inversion recovery (STAIR) prepared UTE (STAIR-UTE) sequence.
- To suppress water signals and enable imaging of ultrashort T2 protons of myelin in white matter using a clinical 3-T scanner.
Main Methods:
- An optimization framework determined the optimal inversion time for nulling water signals using STAIR-UTE at various TRs.
- Numeric simulations, phantom studies, and MRI scans of healthy volunteers and multiple sclerosis (MS) patients were conducted.
- Comparison with a clinical T2-weighted fluid-attenuated inversion recovery sequence for MS lesion assessment.
Main Results:
- STAIR-UTE effectively suppressed long T2 tissues in simulations and phantom studies.
- The 3D STAIR-UTE sequence successfully suppressed water components and imaged myelin (T2* = 0.21 msec ± 0.04) in volunteers.
- Significantly lower myelin proton density was observed in MS lesions and normal-appearing white matter (NAWM) compared to normal white matter (NWM).
Conclusions:
- The STAIR-UTE sequence provides efficient water signal suppression for volumetric myelin imaging.
- It offers excellent myelin signal contrast and demonstrates significant ultrashort echo time signal reduction in MS lesions and NAWM.
- This technique shows promise for assessing myelin integrity in neurological conditions like MS.