Related Experiment Video
Updated: Dec 12, 2025

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Biomechanics of external fixator of distal radius fracture, a new approach: Mutifix Wrist
N Della Rosa1, N Bertozzi2, R Adani3
1O.U. of Hand Surgery and Microsurgery; Azienda Ospedaliera Policlinico of Modena, Largo del Pozzo, 1, 41100, Modena, MO, Italy. normandellarosa@yahoo.it.
Background:
The purpose of this study was to assess factors affecting fixator stiffness with a finite element model and an experimental validation with particular attention to a new fixator device named Mutifix Wrist® (NCS Lab srl, Carpi, MO, Italy).
Materials And Method:
Mechanical tests were carried out to determine the stiffness of the construct with different configurations. The obtained results were compared to those obtained with the Hoffmann II Compact (Howmedica-Osteonics Inc. Rutherford, NJ, USA). Data were sampled at 20 Hz and test speed was 0.05 mm/s. For each loading condition, tests were performed four times. A FEM campaign was also conducted to analyze how geometrical variables (type of configuration and K-wire diameter) affect both stiffness and stress distribution of the fixator.
Results:
Stiffness, axial displacement, magnitude of the displacement, magnitude and localization of the peak stress of the construct were all analyzed. Axial compression tests showed that the axial displacement reached by the machine actuator when the measured force reached 45 N was 0.56 mm ± 0.13 on average. Magnitude of the displacement along with peak stress magnitude and localization varied through the several configuration tested, but it resulted that the distal pin near the fracture gap was the more stressed one in all cases except those in which the fracture line is crossed.
Conclusions:
From the tests performed, it emerged that the addiction of a K-wire provides a construct stiffening and a consequent local stress reduction; while span increase reduces stiffness and increase the local stress. If a K-wire is implanted through the fracture site, the axial stiffness is significantly increased.
More Related Videos
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
05:34A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019
Related Concept Videos
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...
Bones of the Upper Limb: Ulna
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Muscles of the Forearm that Move the Hand and Fingers
Anterior Compartment
The anterior compartment muscles originate from the humerus. They primarily function as flexors and are also known as flexor muscles. They typically insert on the carpals, metacarpals, and phalanges. The superficial layer includes the flexor carpi...