First-stage validation of a portable imageable MR-compatible water calorimeter
Mark D'Souza1, Humza Nusrat2,3, James Renaud4,5
1Department of Physics, Ryerson University, 350 Victoria St., Toronto, ON, M5B 2K3, Canada.
Purpose:
The purpose of this study is to design a water calorimeter with three goals in mind: (a) To be fully magnetic resonance (MR)-compatible; (b) To be imaged using kV cone beam computed tomography (CBCT), MV portal imaging or MRI for accurate positioning; (c) To accommodate both vertical and horizontal beam incidence, as well as volumetric deliveries or Gamma Knife®. Following this, the calorimeter performance will be measured using an accelerator-based high-energy photon beam.
Methods:
A portable 4°C cooled stagnant water calorimeter was built using MR-compatible materials. The walls consist of layers of acrylic plastic, aerogel-based material acting as thermal insulation, as well as tubing for coolant to flow to keep the calorimeter temperature stable at 4°C. The lid contains additional pathways for coolant to flow through as well as two hydraulically driven stirrers. The water calorimeter was positioned in an Elekta Versa using kV CBCT imaging as well as orthogonal MV image pairs. Absolute absorbed dose to water was then determined under a 6 MV flattening filter-free (FFF) beam. This was compared against reference dosimetry results that were measured under identical conditions with an Exradin A1SL ionization chamber with a calibration coefficient directly traceable to the National Research Council Canada.
Results:
The dose to water determined with the calorimeter (n = 30) agreed with the A1SL ionization chamber reference dose measurements (n = 15) to within 0.25%. The uncertainty associated with the water calorimeter absorbed dose measurement was estimated to be 0.54% (k = 1).
Conclusions:
An MR-compatible water calorimeter was successfully built and absolute absorbed dose to water under a conventional 6 MV FFF beam was determined successfully as a first-stage validation of the system.
Related Concept Videos
Constant Pressure Calorimetry
Calorimetry
Constant Volume Calorimetry
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...