The Haloscale "Infanta" Wright respirometer. An in vitro and in vivo assessment

D J Hatch1, G M Williams

  • 1Hospital for Sick Children, London.

Insights

The Haloscale Infanta respirometer accurately measures volumes between 15-200 ml. Performance decreased outside this range, showing potential limitations for critical respiratory monitoring.

Area of Science:

  • Pediatric respiratory monitoring
  • Medical device performance evaluation

Background:

  • Accurate measurement of respiratory volumes is crucial for pediatric patient care.
  • Existing respirometry devices may have limitations in accuracy and range.

Purpose of the Study:

  • To evaluate the performance of the Haloscale Infanta respirometer.
  • To determine the accuracy and limitations of the Infanta respirometer across different respiratory conditions.

Main Methods:

  • In vitro testing using ISO standards for compliance and resistance.
  • In vivo assessment comparing Infanta volumes to pneumotachograph data in children.
  • Testing included both spontaneous breathing and intermittent positive pressure ventilation.

Main Results:

  • The Infanta respirometer demonstrated an accuracy of +/- 5% for volumes between 15 and 200 ml.
  • Volume registration accuracy rapidly declined for volumes below 15 ml.
  • Over-registration of volumes was observed above 200 ml.

Conclusions:

  • The Haloscale Infanta respirometer is accurate for tidal volumes within the 15-200 ml range.
  • Clinical use should consider the device's limitations at very low (<15 ml) and high (>200 ml) volumes.
  • Further validation may be needed for specific pediatric respiratory applications.