Related Experiment Video
Updated: Dec 12, 2025

Author Spotlight: Investigating Islet Abnormalities and Function with a Pseudoislet Protocol
Published on: November 3, 2023
A micro-bioimpedance meter for monitoring insulin bioavailability in personalized diabetes therapy
Pasquale Arpaia1,2, Umberto Cesaro3, Mirco Frosolone4
1CIRMIS - Interdepartmental Center for Research in Management and Innovation in Healthcare, University of Naples Federico II, Naples, Italy. pasquale.arpaia@unina.it.
Abstract:
An on-chip transducer, for monitoring noninvasively the insulin bio-availability in real time after administration in clinical diabetology, is proposed. The bioavailability is assessed as insulin decrease in situ after administration by means of local impedance measurement. Inter-and-intra individual reproducibility is enhanced by a personalized model, specific for the subject, identified and validated during each insulin administration. Such a real-time noninvasive bioavailability measurement allows to increase the accuracy of insulin bolus administration, by attenuating drawbacks of glycemic swings significantly. In the first part of this paper, the concept, the architecture, and the operation of the transducer, as well as details about its prototype, are illustrated. Then, the metrological characterization and validation are reported in laboratory, in vitro on eggplants, ex vivo on pig abdominal non-perfused muscle, and in vivo on a human subject, using injection as a reference subcutaneous delivery of insulin. Results of significant intra-individual reproducibility in vitro and ex vivo point out noteworthy scenarios for assessing insulin bioavailability in clinical diabetology.