Compliance of Textile Vascular Prostheses Is a Fleeting Reality
Benoit Lucereau1, Foued Koffhi2, Anne Lejay1
1GEPROVAS, Groupe Européen de Recherche sur les Prothèses appliquées à la Chirurgie Vasculaire, Strasbourg, France; Department of Vascular Surgery and Kidney Transplantation, University Hospital of Strasbourg, Strasbourg, France.
Objective:
Compliance is considered to be a major property influencing the long term performances of synthetic vascular substitutes that could play a role in anastomotic false aneurysm and intimal hyperplasia stenosis onset. Over the last decades, manufacturers have tried to develop substitutes that mechanically mimic arterial properties and avoid a compliance mismatch at the anastomoses in particular. However, data are missing about how initial compliance properties could change with time. The goal of this study was to evaluate how the compliance of vascular grafts evolves under cyclic loading conditions in vitro.
Methods:
The compliance of three different models of commercially available textile polyethylene terephthalate (PET) grafts was evaluated. Tests were performed with and without their original coating. Compliance was assessed with a specific device dedicated to measure the deformations undergone by a graft under cyclic pressure loading conditions, using image analysis software. In each experiment, image analysis was performed under 60 and 140 mmHg pressure loading conditions at loading start (H0) and after three, six, and 24 h (H3, H6, H24) loading time. Average radial, longitudinal, and volumetric compliance was calculated from the obtained images.
Results:
Twenty-four samples were tested. Results demonstrate that all values decreased significantly within only a few hours. On average, the loss of compliance after 3 h of cyclic loading ranged on average from 35% for longitudinal compliance to 39% for radial compliance and 37% (p < .050) for volume compliance. After 24 h, the loss of radial, longitudinal and volume compliance was respectively 63 ± 3%, 60.5 ± 2% and 61 ± 7%.
Conclusion:
In this in vitro model, PET graft compliance has already decreased significantly within 3 h. The rapid loss of compliance identified in this experimental study helps explain the mismatch mentioned in clinical observations.
More Related Videos
11:49Autologous Endothelial Progenitor Cell-Seeding Technology and Biocompatibility Testing For Cardiovascular Devices in Large Animal Model
Published on: September 9, 2011
11:17Surgical Technique for the Implantation of Tissue Engineered Vascular Grafts and Subsequent In Vivo Monitoring
Published on: April 3, 2015
Related Concept Videos
Peripheral Artery Disease V: Postoperative Nursing Management
Hemodialysis I: Introduction