Repeatability of Zone Averages Compared to Single-Point Measurements of Maximal Curvature in Keratoconus
Lara Asroui1, Carl-Joe Mehanna1, Ali Salloum1
1Department of Ophthalmology, American University of Beirut Medical Center, Beirut, Lebanon.
Purpose:
To evaluate the repeatability of curvature zone averages centered on the point of maximum curvature (Kmax) compared to that of the single-point Kmax.
Design:
Comparative reliability analysis.
Methods:
Setting: American University of Beirut Medical Center, Beirut, Lebanon.
Study Population:
Sixty-five eyes of 65 adult keratoconus patients. Patients with other ocular disease, history of ocular surgery or trauma, and contact lens wear within 2 weeks of image acquisition were excluded.
Observation Procedures:
Eyes were evaluated with 3 consecutive scans using the Galilei dual Scheimpflug-Placido system.
Main Outcome Measures:
Repeatability of axial and instantaneous Kmax single points, and zone averages with radii of 0.1-2.0 mm, centered on them. Repeatability was assessed by within-subject standard deviations, repeatability limits (r), and intraclass correlation coefficients.
Results:
Axial curvature zone averaging yielded clinically acceptable repeatability only in eyes with Kmax ≤50 diopters (D), for radii of 1.5 mm and 2.0 mm (r = 0.87 D and r = 0.76, respectively, vs r = 0.91 for the single-point axial Kmax). Compared to instantaneous Kmax, clinically acceptable repeatability was achieved with instantaneous zone averages of at least 1.5 mm radius in eyes with Kmax ≤50 D (r = 0.99 D and r = 0.70 D, respectively) and 2.0 mm radius in eyes with Kmax >50 D (r = 2.28 D and r = 0.87 D, respectively). For all eyes, the repeatability limit of the location of Kmax was 0.82 mm and 0.80 mm for axial and instantaneous curvature, respectively.
Conclusions:
Instantaneous curvature zone averages centered on Kmax yielded a greater improvement in repeatability than axial zone averages and reached clinical adequacy with radii of at least 1.5 mm, for eyes with Kmax ≤50 D, and with a 2.0 mm radius for eyes with Kmax >50 D.