Effect of image quality fluctuations on the repeatability of thickness measurements in swept-source optical coherence
Heon Yang1, Hye Sun Lee2, Hyoung Won Bae3
1Kong Eye Center, Seoul, Republic of Korea.
Abstract:
This study investigated the effect of image quality fluctuations on the repeatability of thickness measurements of the peripapillary retinal nerve fibre (PP-RNFL) and ganglion cell-inner plexiform (GC-IPL) layers using swept-source optical coherence tomography (SS-OCT). Three consecutive OCT scans each were performed on 56 healthy subject. Finally, 168 SS-OCT results were analysed. Based on the tertile values of the mean absolute difference of image quality score, all subjects were divided into the following three groups-low-(LIQD), moderate-(MIQD), and high-(HIQD) image quality score difference groups. A linear mixed model and intraclass correlation coefficients (ICCs) were used for analyses. Despite high ICC values (> 0.9), several sectors showed significant differences in the ICC values in intergroup comparisons. For LIQD-HIQD and MIQD-HIQD, most PP-RNFL sectors showed significant differences. For GC-IPL sectors, the LIQD-HIQD comparison showed significant differences in the temporosuperior (p = 0.012), inferior (p < .001), and temporoinferior (p = 0.042) sectors. Significant differences existed in the average GC-IPL (p = 0.009), nasoinferior (p = 0.035), and inferior GC-IPL sectors (p < .001) for MIQD-HIQD comparison. With higher image quality fluctuations, the repeatability of SS-OCT decreased in several sectors, which are considered clinically relevant in evaluating glaucoma status. Therefore, maintaining high-quality image status is essential to enhance the reliability of SS-OCT.
More Related Videos
09:56Automated 3D Optical Coherence Tomography to Elucidate Biofilm Morphogenesis Over Large Spatial Scales
Published on: August 21, 2019
04:51Author Spotlight: Characterizing Environmental Biofilm Mechanics Using Optical Coherence Elastography and its Applications in Wastewater Treatment
Published on: March 1, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography