Cardiac Magnetic Resonance Imaging at 7 Tesla
Published on: January 6, 2019
3.0 T magnetic resonance imaging scanning on different body regions in patients with pacemakers
Xiaohui Ning1, Xiaofei Li1, Xiaohan Fan2
1Department of Arrhythmia, State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, A 167 Bei Li Shi Road, Xicheng District, Beijing, 100037, China.
Purpose:
Magnetic resonance imaging (MRI) at 3.0 T is becoming more common, but there is a lack of sufficient evidence on the safety of a 3.0 T scan in patients with pacemakers. This study aimed to investigate the safety and practical concerns of 3.0 T scans for patients with MR-conditional pacemakers.
Methods:
Twenty consecutive patients were enrolled. A standardized protocol was developed by cardiologists, pacemaker engineers, and radiologists. Pacemaker interrogation was performed immediately before and after the scan. Scan-related adverse events were documented, and imaging quality was graded as level 1 to 4 by radiologists.
Results:
Twenty-three MRI scans of different body regions (brain = 13, lumbar spine = 4, cervical spine = 2, and heart = 4) were performed, and the average time of a scan was 25 ± 11 min. No significant changes in sensing amplitude (atrial 3.1 ± 1.1 mV vs. 2.9 ± 1.2 mV, P = 0.71; ventricular 9.3 ± 3.5 mV vs. 10.2 ± 3.4 mV, P = 0.46), lead impedances (atrial 647 ± 146 Ω vs. 627 ± 151 Ω, P = 0.7; ventricular: 780 ± 247 Ω vs.711 ± 226 Ω, P = 0.36), or pacing threshold (atrial 0.6 ± 0.2 V/0.4 ms vs. 0.6 ± 0.2 V/0.4 ms, P = 0.71; ventricular 0.7 ± 0.3 V/0.4 ms vs. 0.7 ± 0.2 V/0.4 ms, P = 0.85) were observed pre- and postscan. No adverse events were detected. Image quality review showed grade 1 quality in 16 patients and grade 2 quality in 4 patients with artifacts of pulse generators and leads in cardiac MRI scan and no impact on diagnostic value.
Conclusion:
Our initial data indicated that 3.0 T scanning might be feasible under a standardized protocol with good diagnostic imaging quality irrespective of body region in patients with MR-conditional pacemakers.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
12:29Magnetic Resonance Imaging Quantification of Pulmonary Perfusion using Calibrated Arterial Spin Labeling
Published on: May 30, 2011
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...