Comparison of PI-RADS Versions 2.0 and 2.1 for MRI-based Calculation of the Prostate Volume

Soleen Ghafoor1, Anton S Becker1, Sungmin Woo1

  • 1Department of Radiology, Memorial Sloan Kettering Cancer Center, 1275 York Avenue, New York, NY 10065, USA.

Academic Radiology
|August 21, 2020
PubMed
Abstract

Insights

Prostate gland volume (PGV) measurements using PI-RADS v2.0 with the ellipsoid formula are most accurate. The bullet formula significantly overestimates PGV, and PI-RADS v2.1 changes show minor but potentially relevant differences.

Area of Science:

  • Radiology
  • Medical Imaging
  • Urology

Background:

  • Prostate gland volume (PGV) is crucial for prostate MRI reports.
  • The Prostate Imaging Reporting and Data System (PI-RADS) v2.1 updated PGV measurement methods compared to v2.0.

Purpose of the Study:

  • Evaluate MRI-based PGV calculation agreement with manual segmentation.
  • Compare PI-RADS v2.0 and v2.1 measurement methods.
  • Assess inter-reader agreement and the impact of transition zone enlargement.

Main Methods:

  • Retrospective analysis of 95 treatment-naive patients undergoing prostate MRI.
  • Manual prostate segmentation and 3D measurements by four radiologists using PI-RADS v2.0 and v2.1.
  • PGV calculation via ellipsoid and bullet formulas, compared to reference standard.

Main Results:

  • Excellent inter-reader agreement for both ellipsoid and bullet formulas (ICC > 0.985).
  • PI-RADS v2.0 ellipsoid formula showed minimal median difference (0.4 mL) from reference.
  • Bullet formula significantly overestimated PGV (median difference > 13 mL).

Conclusions:

  • PI-RADS v2.0 measurements using the ellipsoid formula provide the most accurate PGV estimates.
  • The bullet formula should be avoided for prostate volume estimation due to overestimation.
  • PI-RADS v2.1 differences are statistically significant and may impact clinical applications like PSA density.