Determination of the cross-contamination and validation of the cleaning process for an automated personalised dosing system
Related Experiment Video
Updated: Dec 11, 2025

Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Alexander Ferro Uriguen1, Idoia Beobide Telleria2, Silvia Martínez Arrechea2
1Pharmacy Service, Hospital Ricardo Bermingham, San Sebastian, País Vasco, Spain alex.ferro@matiafundazioa.eus.
Objective:
To determine the degree of cross-contamination and to validate a cleaning process for an Automated Personalised Dosing System (APDS), respecting the permitted residue transfer limits.
Method:
An analytical determination by high pressure liquid chromatography with mass spectrometry was performed for 11 drugs considered critical for their potential for contamination, toxicity (National Institute for Occupational Safety and Health (NIOSH) List 2016) and/or narrow therapeutic index. The test was carried out for three consecutive weeks, seven repetitions per week, in a state of maximum contamination and maximum cleanliness. The main validation criterion of the cleaning protocol was to quantify that the active ingredients measured were below the Permitted Daily Exposure index or fraction 1/1000 of the minimum daily dose, both in a state of maximum contamination and when the APDS is in a clean condition, for an average number of tablets consumed per person per day.
Results:
In all the samples analysed, some compound could always be quantified below 2% of the permitted transfer limits, highlighting the presence of paracetamol and carbamazepine. In the state of maximum contamination, the mean±SD values of paracetamol and carbamazepine were 5.83±1.56 µg and 0.22±0.07 µg, respectively, without significant differences in mean values over the 3 weeks of the study. After the cleaning protocol was executed, only paracetamol and carbamazepine were detected, with the average values being 4.67 µg (95% CI 3.92 to 5.43, p<0.05) and 0.07 µg (95% CI 0.03 to 0.10, p<0.05) lower than the state of maximum contamination, respectively.
Conclusions:
In all cases the level of cross-contamination for an APDS was below the pre-established limits. The cleaning protocol has been validated, confirming APDS decontamination of the most critical medicines.
More Related Videos
08:29An Open Source Technology Platform to Manufacture Hydrogel-Based 3D Culture Models in an Automated and Standardized Fashion
Published on: March 31, 2022
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Related Concept Videos
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Data Validation
Key parameters for method validation include:
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...