Assessment of S-NPP VIIRS On-orbit Radiometric Calibration and Performance.
Xiaoxiong Xiong1, James Butler1, Kwofu Chiang2
1Sciences and Exploration Directorate, NASA/GSFC, Greenbelt, MD, 20771 USA.
Summary
The Visible Infrared Imaging Radiometer Suite (VIIRS) sensor has exceeded four years of operation, providing high-quality environmental data. Its on-orbit calibration ensures reliable performance, meeting all design requirements despite some degradation.
Area of Science:
- Earth Observation
- Remote Sensing Technology
- Spacecraft Instrumentation
Background:
- The Visible Infrared Imaging Radiometer Suite (VIIRS) instrument on the Suomi-NPP spacecraft has been operational since October 2011.
- VIIRS generates environmental data records (EDR) from sensor data records (SDR) for operational and research applications.
Purpose of the Study:
- To review VIIRS sensor on-orbit calibration methodologies for reflective solar bands (RSB) and thermal emissive bands (TEB).
- To assess the on-orbit radiometric performance of VIIRS.
- To describe improvements in calibration and data quality.
Main Methods:
- Utilized measurements from instrument on-board calibrators (OBC) and lunar observations for radiometric performance assessment.
- Controlled temperatures of the on-board blackbody (BB) and cold focal plane assemblies for stability.
- Monitored and analyzed degradation in optical throughput and solar diffuser.
Main Results:
- All on-board calibrators (OBC) operated normally, enabling critical calibration parameter updates.
- Stable temperature control was maintained for the on-board blackbody (BB) and focal plane assemblies.
- Despite degradation in infrared bands and solar diffuser, VIIRS performance met design requirements.
Conclusions:
- VIIRS demonstrates robust radiometric performance and data quality, meeting mission objectives.
- Ongoing efforts focus on addressing identified challenges to further enhance sensor calibration and characterization.
- Continuous calibration and quality improvements ensure the long-term utility of VIIRS data for Earth science applications.
More Related Videos
Related Concept Videos
Instrument Calibration
579
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
579
Calibration Curves: Correlation Coefficient
4.4K
In a linear calibration curve, there is a value called the calibration coefficient, denoted by 'r,' which measures the strength and the direction of association between two variables. The correlation coefficient value ranges from −1 to +1. A value of +1 indicates a perfect positive linear correlation, −1 denotes a perfect negative correlation, and 0 implies no correlation between the two variables. A positive correlation value establishes that as one variable increases, the...
4.4K
Uncertainty in Measurement: Reading Instruments
49.5K
Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
49.5K
Glassware Calibration
1.2K
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
1.2K